/ Instrumentation


VME-based data acquisition system for multiparameter measurements


Slepnev, R.S. (Dubna, JINR) ; Bezbakh, A.A. (Dubna, JINR) ; Golovkov, M.S. (Dubna, JINR) ; Gorshkov, A.V. (Dubna, JINR) ; Gorshkov, V.A. (Dubna, JINR) ; Daniel, A.V. (Dubna, JINR) ; Kaminski, G. (Cracow, INP) ; Krupko, S.A. (Dubna, JINR) ; Martianov, A.S. (Dubna, JINR) ; Sidorchuk, S.I. (Dubna, JINR) ; Chudoba, V. (Silesian U., Czech Republic) ; Fomichev, A.S. (Dubna, JINR)

Published in: Instrum.Exp.Tech. / Prib.Tekh.Eksp.
Year: 2012 / 2012
Vol.: 55 / 6
Page No: 645-650 / 41-46
Pages: 6

Abstract: The VME-based data acquisition system designed for physics experiments involving measurements of several hundreds of parameters is described. The equipment has been tested in the experiments with the (6)He, (8)He, (18)Ne, and (27)S beams at the ACCULINNA fragment separator ( http://aculina.jinr.ru/ ) in the energy range of 20-35 MeV/nucleon. The system is composed of the RIO-3 processor module, which is used to read the VME bus, communicate with the CAMAC and FASTBUS crates, and transmit data to the personal computer, the TRIVA-5 master trigger module, standard analog-to-digital, charge-to-digital, and time-to-digital converters from CAEN (V785, V792, and V775 models), and related software. This data acquisition system is faster (by a factor of 10) than a CAMAC-based system and is capable of operating with new types of electronic modules, e.g., with digitizers.

Total numbers of views: 2099
Numbers of unique views: 625
DOI: 10.1134/S0020441212050065
 Record created 2013-01-10, last modified 2014-01-30



Rate this document:

Rate this document:
1
2
3
 
(Not yet reviewed)